Wellsystem medical Gebrauchsanweisung Seite 121

Vorschau ausblenden Andere Handbücher für medical:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Interference immunity tests
Discharging static electricity
(ESD)
IEC 61000-4-2
Fast transient electrical noise /
bursts
IEC 61000-4-4
Surges
61000-4-5
Magnetic field radiated by supply
frequency (50/60 Hz)
IEC 61000-4-8
Voltage drops,
Short interruptions
and supply voltage
fluctuations
IEC 61000-4-11
10/2017
1011906-02A
7.1.2
Table 202: Guidelines and manufacturer's
declaration – Electromagnetic interference
The wellsystem™ MEDICAL massage device has been designed for use in
the electromagnetic environment specified below. The device operator must
ensure that it is used such an environment.
IEC 60601 - Test level
±6 kV contact discharge
±8 kV air discharge
±2 kV for mains supplies
±1 kV series mode voltage
±2 kV common mode voltage
3 A/m
<5% U
(>95% drop in U
)
T
T
for 0.5 periods
40% U
(60% drop in U
)
T
T
for 5 periods
70% U
(30% drop in U
)
T
T
for 25 periods
<5% U
(>95% drop in U
)
T
T
for 5 s
Note: U
is the mains alternating voltage used for the test levels.
T
Defined level
±6 kV contact discharge
±8 kV air discharge
±2 kV for mains supplies
±1 kV series mode voltage
±2 kV common mode voltage
3 A/m
<5% U
(>95% drop in U
)
T
T
for 0.5 periods
40% U
(60% drop in U
)
T
T
for 5 periods
70% U
(30% drop in U
)
T
T
for 25 periods
<5% U
(>95% drop in U
)
T
T
for 5 s
Operating instructions – 55/62
Appendix
Electromagnetic
environment –
Guidelines