Replacement of the nebulizer
The nebulizer must be replaced after a
long period of inactivity, or when it is
obstructed by dry medication, dust, etc.
Your nebulizer should last from 6 months
to 1 year with general use.
Use original nebulizers only. (Year Set Pro
nebulizer M 53710-04)
Replacement of the filter
The MicroDrop
®
Pro Porex filter must be
replaced approximately after 500 hours of
use or each year; control the filter periodi-
cally (10 - 12 aerosol treatments); when it
turns grey or brown or wet, replace it;
open the filter holder (pic. A / 5) and replace
the filter with a new one. Don't try to clean
the filter and to reuse. Use original filters
only.
Do not use the device without filter.
Possible Problems and how
to solve them
If the device does not switch on:
1. make sure the plug is firmly fitted to
the wall socket;
2. the thermal protector might have been
released. The thermal protector may be
released for the following reasons:
– the device has been in operation for a
too long time;
– the device was used near heaters or in
hot environments;
24
– the air vent openings (pic. A / 7) are
obstructed and do not allow for the
motor ventilation.
In this case, set the ON/OFF switch to
"O" and let it rest for at least 15 minutes.
If the device does not nebulize, or only
very weakly, check the following points:
1. make sure that the ends of the air tube
are fitted tightly onto the main unit and
onto the nebulizer;
2. check whether the nebulizer is empty
from medication or full of an excessive
medication quantity (max. 15 ml);
3. check whether the nebulizer nozzle is
obstructed;
4. check whether the air conduction cone
is inserted (pic. E).
Should the device not start
working properly again,
address to an authorized technical
service centre.

Maintenance and Repairs

In case of failure, address to qualified
personnel authorized by MPV TRUMA.
Never open the device.
There are no user-serviceable parts within.
The compressor requires no lubrication or
maintenance.