The capacity of a fully charged battery is sufficient for one full day.
We recommend charging the product overnight when used by the patient on a daily basis.
When used daily, the complete charging unit (power supply – battery charger) may remain
plugged into the wall socket.
The battery should be charged for at least 3 hours prior to initial use.

6.1 Connecting the power supply and battery charger

1) Slide the country-specific plug adapter onto the power supply until it locks into place.
2) Connect the round, three-pin plug of the power supply to the 12V receptacle on the battery
charger so that the plug locks into place.
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting
the cable plug to the battery charger.
3) Connect the round, four-pin plug of the charging cable to the OUT receptacle on the battery
charger so that the plug locks into place.
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting
the cable plug to the battery charger.
4) Plug the power supply unit into the wall socket.
→ The green LED on the back of the power supply and the green LED on the battery charger
light up.
→ If the green LED on the power supply and the green LED on the battery charger do not light
up, there is an error (see Page 65).

6.2 Connect battery charger to the product

6.3 Display of the current charge level

The current battery charge level can only be displayed by connecting the battery charger and
power supply to the product.
50 | Ottobock
1) Open the charging receptacle cover.
2) Connect the charging plug to the charging receptacle of the
product.
INFORMATION: Take note of the plug direction!
→ The correct connection of the battery charger to the product is
indicated by feedback (Status signals).
3) The charging process starts.
→ After approx. 6 hours, the product battery is fully charged and
the yellow LED on the battery charger turns off.
4) Disconnect the product after the charging process is complete.
→ This is followed by an electronics self-test which is confirmed by
feedback (Status signals).
5) Close the charging receptacle cover.