Please follow the step by step guide.
General safety precautions:
Highest safety precautions should be used when dispensing corrosive, poisonous, radioactive or hazardous
chemicals.
Observe general safety regulations (e.g. wear safety goggles, protective clothing, etc).
Carefully follow the operation manual.
Use the instrument only with regard to resistance of material and for the purpose for which it is intended. In case
of doubt concerning the suitability please contact the manufacturer.
Always check the dispenser for leaks. Make sure that the suction and ejection canula are firmly in position before
starting to use the instrument.
NEVER use force! Breakage of any part will lead to hazardous exposition to the user/ operator as well as others.
We only consider ourselves responsible in regard to safety, reliability and performance of the MINISPENSOR if:
The dispenser has been used in accordance to the instruction manual
Only the parts supplied by the manufacturer have been used
Before operating the dispenser ...
Check if dispenser has not been damaged during transport.
The Minispensor is suitable for practically all liquids and dispensing methods.
EXCEPTION:
Solutions of hydrofluoric acid or of substances which form crystals, contain or form solids, as well as solutions of
substances which are catalyzed by platinum and iridiuma
reagent) may precipitate metallic oxides, therefore clean the dispenser daily:
The temperature of the instrument and the reagent should not exceed 40 °C (104 °F).
The user has to ensure the compatibility of the dispenser and its chemical resistance for himself.
Select the appropriate length of the canula according to the bottle which will be used and push it on tightly. Screw the
ejection canula on firmly.
Do not use the plunger before the dispenser has been assembled completely!
The Minispensor fits to the corresponding 100 ml bottle, also supplied with an adapter A28/32; more adapters, see
SPARE PARTS LIST
Warning: Avoid personal injury from chemicals. To protect yourself always wear eye protection and other
appropriate safety equipment and clothing. Please follow all safety instructions and observe the operating procedures.
Safe and proper operation and function is only possible with the ejection canula as originally supplied. Never use
damaged or deformed canulas. Do not use any other type of tubing. Regularly inspect the tubes for leakage and
replace immediately damaged or deformed ones.
Operation Manual
MINISPENSOR
7
alloy. Inorganic oxidizable solutions (e.g. biuret