Starting and Stopping a Run

3.5.
Please refer Chapter 5 Operation to set up your preferred settings of time, speed and temperature.
To start a run, press down the
When the pre-set time expires, the centrifuge will stop automatically. To stop the centrifuge prior to
the expiration of set time, press the
When the time is set as "continuous"( - - ), the centrifuge will run continuously. To stop the
continuous run, press the
By pressing and holding the
short-run will stop if the
Note:
1. For safety reasons, the
symbol "
" will be flashing until the lid is closed.

Emergency Lid Release

3.6.
When the lid can't be opened by pressing the
off the power of the centrifuge. Second, use a flat-head screw driver or other tool to remove the
Emergency Lid Release Plug ( Please refer Figure 2 ) on the right-hand side of the machine. Third,
pull the wire (attached to the plug) while gently pushing downward on the lid. The lid will now
release.
4
Figure 1. Loading the rotor
(START/STOP) knob.
(START/STOP) knob again.
(START/STOP) knob.
button, the centrifuge will continue a short-run of set speed. The
button is released.
(START/STOP) knob will be inactivated when the lid is open. A
button, users can open the lid manually. First, turn