MAINTENANCE MANUAL for the
1.
Remove the mast from the
base (as per §III.2.4) to be
able to work on the spread-
er bar
2.
Loosen the lock nut of
the tie-rods using the flat
spanner
3.
Remove the head of the
tie-rods, and ball joints on
the pedal
4.
Loosen the screw (inside
the spreader bar) and nut
(below the spreader bar)
securing the pedal.
Use the bushing and the flat
spanner
5.
Extract the pedal assembly
from the spreader bar
Samsoft Elite 18O
EN
PATIENT LIFT
43