Speed control knob: adjust the speed, clockwise value increases, counterclockwise value decreases; press
this key to switch to the timing mode, and the value increases clockwise and decreases counterclockwise.
3.
Operation instructions
1)
Basic settings
A.
Inching mode:
Press the switch to activate inching mode; the indicator light on the panel will turn on, and the speed
is not adjustable.
Place the sample tube for mixing at the center of the standard head, ensuring the tube is vertical.
Gently press the sample tube to initiate the oscillation of the standard head.
After the mixing is completed, remove the sample tube from the standard head to stop the operation.
B.
Continuous operation mode:
Replace the standard head with a universal fixture and tighten the screws.
Insert the centrifugal tubes with mixed samples evenly into the test tube module, ensuring the sample
volumes in each tube are approximately equal. Place the test tube module into the universal fixture
and secure it.
Press the switch to activate continuous operation mode; the continuous operation light will turn on,
and speed adjustment is not available.
The product begins operation. After completion, turn the switch to the center position to stop the
operation.
2)
Speed control mode settings
A.
Inching mode:
Press the "Touch/Cont." button to switch to inching mode; the indicator will light up, and the speed is
adjustable.
Rotate the speed control knob counterclockwise to set the minimum speed.
Place the sample tube for mixing at the center of the standard head, keeping the tube vertical.
Gently press the sample tube and gradually rotate the speed control knob to adjust to the desired
speed. The standard head will begin oscillating.
After mixing, remove the sample tube from the standard head to stop the operation.
B.
Continuous operation mode:
Replace the standard head with the universal fixture and securely tighten the screws.
Insert the centrifuge tubes with mixed samples evenly into the test tube fixture, ensuring that the
sample volumes in each tube are approximately equal. Place the test tube fixture into the universal
fixture and fasten it securely.
Rotate the speed control knob counterclockwise to the minimum speed setting.
Press the "Touch/Cont." button to switch to continuous operation mode. The continuous operation
light will turn on, and speed adjustment becomes available.
Slowly rotate the speed control knob to set the desired speed, and the product will begin running.
EN