Pull a single adhesive backed replica tape free from its release
paper, being careful not to touch either side of the replica foam
during application.
4a. For flat or convex surfaces: Using a
Testex High Accuracy Burnishing
Tool, press the ball of the tool onto
the replica foam with just enough
pressure so that the flat base rests on
the tape. Move the tool's ball around
the circular replica foam opening.
For concave surfaces: Flip over the High Accuracy Burnishing
Tool and hold it at an angle. Use the curved area near the Testex
logo at the top of the tool to burnish.
4b. If using a Classic Burnishing Tool,
hold the tool like a pencil, and apply
consistent pressure with the ball end,
being careful not to apply too much
force. Excessive force can lead to
inaccurate measurements.
NOTE: The stick-type legacy burnishing tool remains a valid
measurement practice and is recognized by international
standards, but does not have an associated accuracy statement.
See Accuracy (pg. 5).
3. Apply the replica tape to the surface.
Press down on the adhesive backing of
the tape around the replica foam to
ensure it remains in place during the
burnishing process.
— or —