When Conversion Mode is disabled, an H icon
is displayed at
the bottom center of the measurement screen. The gage displays the
measured thickness without adjustment, comparable to an analog
spring micrometer. A correction table must be used to accurately
measure surface profile using replica tape.
NOTE: Conversion Mode is unavailable when No Replica Tape Grade
is selected from the gage menu.
X2 X2
Averaging Mode
At least two measurements should be taken at each location. If the two
replicas differ by 5 µm (0.2 mils) or less, their average should be
recorded as the profile height.
X2 X2
When Averaging Mode
is enabled, the gage will display the
average of two measurements. The individual readings will be
displayed as R1 and R2, and the average of the two readings will be
displayed on the main measuring screen.
NOTE: When both Averaging Mode and Memory Mode are ON, only
the average of the two readings is recorded. The number of readings
and statistics will be based on the recorded averages.
NOTE: Averaging Mode is unavailable when No Replica Tape Grade is
selected from the gage menu.
Gage Reset
Restores the gage to default settings, clears stored memory and zero
adjustments.
Power Off
Powers down the gage. Alternatively, the gage will automatically
power down after 5 minutes of no activity. All settings are retained.
Exit Menu
Exits the gage menu and returns to the main measurement screen.
NOTE: While in the menu, pressing down on the lever to open the
anvils will also exit the menu.