Herunterladen Diese Seite drucken

Invacare Easy Fit Gebrauchsanweisung Seite 7

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 13
WARNING!
Risk of serious injury or damage
Excessive force on the handles can cause the sling
to tear.
– Do not use the handles to lift!
4.5 Removing the Sling
1. After completion of the transfer, detach the sling from
the lift.
2. Reverse the procedure described in 4.2 Applying the
Sling, page 6 .
5 Maintenance
5.1 Inspection
Daily Inspection Before Use
Inspection to be performed daily, before use and after each
laundering:
Visually inspect the sling and check all parts for signs of
damage, wear or potential failure.
Periodic Inspection
NOTICE!
Inspections must be performed by a qualified
person who is well acquainted with the design,
use and care of slings.
A periodic safety inspection of the sling must be performed
at least every 6 months unless otherwise stated in local
requirements.
A detailed safety inspection checklist is available from
Invacare as a separate document.
LOLER Statement
The UK Health and Safety Executive's Lifting Operations and
Lifting Equipment Regulations 1998, require any equipment
that is used in the workplace to lift a load be subject to
safety inspection on a six monthly basis. Please refer to the
HSE web site for guidance www.hse.gov.uk.
The person responsible for the equipment must ensure
adherence to LOLER regulations.
5.2 Cleaning and Disinfection
WARNING!
Risk of injury
Usage of damaged slings can cause the patient to
fall or cause injury to assistants.
– After each laundering, inspect the sling for
wear, tear, and loose stitching.
NOTICE!
Regular cleaning and disinfection prevents
contamination.
Clean and disinfect the product
– regularly while it is in use,
– when it has been in contact with any body
fluids,
– before using it for a new user.
NOTICE!
Always make sure that the product is completely
dried before taking it into use again.
Cleaning Instructions
Launder and dry the sling according to the washing
instructions on the label.
60124415-C
Frequent use of high temperatures and tumble drying can
reduce the service life of the product.
We recommend the use of an ordinary household laundry
detergent. Do not use softener.
For disinfenction, it can be done by using a disinfectant on
cleaning procedure. Select a disinfectant that is compatible
with the sling materials and follow the instructions of the
disinfectant manufacturer.
6 After Use
6.1 Storage Conditions
The sling must be stored in dry conditions with a maximum
relative humidity of 50% and without direct sunlight.
6.2 Reconditioning
This product is suitable for reuse. To recondition the product
for a new user, carry out the following actions:
Inspection
Cleaning and disinfection
For detailed information, see 5 Maintenance, page 7 .
Make sure that the user manual is handed over with the
product.
If any damage or malfunction is detected, do not reuse the
product.
6.3 Disposal
Be environmentally responsible and recycle this product
through your recycling facility at its end of life.
Disassemble the product and its components, so the different
materials can be separated and recycled individually.
The disposal and recycling of used products and packaging
must comply with the laws and regulations for waste handling
in each country. Contact your local waste management
company for information.
7