WARNING!
The maximum safe working load must not be
exceeded
– Do not exceed the maximum safe working
load of this product or other components of
the lifting system. See documentation or the
labeling for the stated maximum safe working
load.
– The component with the lowest load limit
determines the maximum safe working load of
the entire system.
WARNING!
Risk of Injury or Damage
Ignition sources can cause burns or fire.
– Patient transfer must be performed with a
safety space between the lift and possible
ignition sources (heater, stove, fireplace, etc.).
– The patient and assistants must not smoke
during transfer.
– The sling must not be placed over heat sources
(heater, stove, fireplace, etc.).
CAUTION!
Compatibility of Slings with Attachment System
Invacare uses a common attachment system
based on hooks and loops. Loops on the sling are
attached to hooks on the lift.
– A risk assessment is always to be carried out
by a professional prior to use of Invacare slings
with lifts of other manufacturers.
– If you are in doubt about the use of Invacare
slings with equipment of other manufacturers,
contact your Invacare provider for advice.
2.2 Labels and Symbols on the Product
The labels are placed on the upper backside of the sling.
Symbols — Fig. 1
A
Read user manual
UK Conformity Assessed
B
European Conformity
C
Manufacturer
D
Serial number
E
Date of manufacture
F
Lot number
G
Unique Device Identifier
H
Reference number
I
Max. safe working load
J
Do not bleach
K
Tumble dry, low temperature
L
Do not iron
M
N
Washing max. 90 °C
Medical device
O
Field for patient name
P
QR code with link to instruction video
Q
60124415-C
3 Product Overview
3.1 Intended Use
The Invacare Easy Fit slings are non-rigid body-support units
intended to be used in combination with a mobile hoist,
stationary hoist fixed to walls, floor or ceiling or stationary
free-standing hoist using a loop to hook attachment with 2
or 4 connection points.
The sling is not intended to be left underneath the patient
after the transfer.
Intended Users
The Invacare Easy Fit sling is intended to a completely or
partially immobile patient.
Indications
The sling is indicated for the transfer of completely or
partially immobile patients with the following minimum
degrees of body control:
Hip control
Trunk control
Limited
Limited
There are no contraindications known when using the
product as intended.
Intended Operator
A healthcare professional or private person who has received
proper training is the intended operator of this product.
3.2 Options
Extension loops – 20 cm or 30 cm length
Intended to be attached to the outer loops and extend
the suspension straps of the sling for more reach or
positioning options.
Head support
Itended to be used for patients that need additional
support that the sling can not provide or occasionally
for a short time if the patient fits best in a lower sling.
Loop marker
Intended to mark the preferred loops after an individual
assessment.
4 Usage
4.1 General Safety Information
WARNING!
Risk of injury or damage
Improper use of this product may cause injury
or damage.
– Do not attempt any transfer without approval
of the patient's healthcare professional.
– Ensure that the patient is not prone to spasm
or a risk assessment has been completed to
address the possibility of forward pitching.
Head control