Herunterladen Inhalt Inhalt Diese Seite drucken

Elma Instruments Elma 3036 Bedienungsanleitung Seite 52

Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Electrical Specifications
Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 23
than 75% R.H.
Models 112M & 128M True RMS accuracy of ACV is specified from 5 % to 100 % of range, and
ACA is specified from 10 % to 100 % of range, or otherwise specified. Maximum Crest Factor <
1.75 : 1 at full scale & < 3.5 : 1 at half scale, and with frequency spectrums, besides
fundamentals, fall within the meter specified AC bandwidth for non-sinusoidal waveforms
DC Voltage
RANGE
400.0 mV
4.000V, 40.00V,
400.0V
600V
NMRR
:
CMRR
:
Input Impedance : 10MΩ, 30pF nominal
(1000MΩ for 400.0mV range)
AC Voltage
RANGE
50Hz -- 500Hz
4.000V, 40.00V,
400.0V
600V
CMRR
:
Input Impedance
ACA Current (Clamp-on)
RANGE
50Hz / 60Hz
40.00A
400.0A
600A**
*Not available to Model 110M
1)
Induced error from adjacent current-carrying conductor: 0.05/A
2)
Specified accuracy is from 1% to 100% of range and for measurements made at the jaw
center. When the conductor is not positioned at the jaw center, position errors introduced are:
Add 2% to specified accuracy for measurements made BEYOND jaw marking lines (toward jaw
opening)
3)
Add 8d to specified accuracy @ reading < 10% of range
Max Hold (where applicable)
Specified accuracy ± 50 digits for changes > 25ms in duration
Audible Continuity Tester
Audible threshold : between 20ohm and 120ohm.
All manuals and user guides at all-guides.com
Elma 3036 + 3038 + BM126
Accuracy
0.3% + 4d
0.5% + 3d
1.0% + 4d
>50dB @ 50/60Hz
>120dB @ DC, 50/60Hz, Rs=1kΩ
1)
Accuracy
1.5% + 5d
2.0% + 5d
>60dB @ DC to 60Hz, Rs=1kΩ
: 10MΩ, 30pF nominal
1)
Accuracy
1.9% + 8d
1.9% + 8d*
unspecified
side 52
O
O
C ±5
C & less
Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Elma 3038Elma 126Bm 111mBm 112mBm 127m

Inhaltsverzeichnis