Specification:
VI-D4 meter was designed in order to measure concrete humidity. Its unquestionable advantage is
nondestructive method of measurement with simultaneous high accuracy and easiness of use.
Features :
Dimension: 147x89x33mm
Power supply: 2xAA battery
Average working time on one battery set : 20 hours
Display: monochrome 128x64 pixels, size 61x33mm with backlight
Operating temperature range: 5
Accuracy: ±0.5%
Scales: -Concrete
-Carbide Method
-Relative Scale
-Caisson Scale
General description:
VI-D4 meter calculates humidity of analyzed material by measuring its electrical impedance. The relation
between humidity of certain material and its impedance is directly proportional. Impedance is measured through
generating low frequency electric field between electrodes. The meter measures low intensity alternating current
flowing through electric field and on that basis calculates moisture of tested material.
Constraints:
Concrete Moisture Meter VI-D4 does not detect concrete humidity through electricity conducting materials like
metal coverings/ linings, PE-rubber or any wet surfaces.
Moisture Meter VI-D4 should not be used to measure concrete humidity on surfaces which are topped with a
thick layer of floor covering material like wood.
Meter measurement outcomes show the actual humidity at a time of measurement.
Concrete Moisture Meter
VI-D4
User's Guide
o
o
C to 40
C