Log function

With the LOG function it is possible to view historical data records.
The datalog is visible by following the steps below:
1. Click
.
2. Click twice
to go to
3. Click
when
4. Click arrow
5. Click arrow
PPM value per time interval.

Calibrate the CO2 monitor manually (CALI)

The CO2 meter can be calibrated manually via the following steps:
1. Leave the CO2 monitor in an open window or outside for at least 20 minutes and make
sure that the CO2 value shown is stable.
2. Click on
and click 3 times on
3. If it displays "400 ppm", the enter key must be pressed for 5 seconds until the
device emits a sound signal.
4. The display will now count down from 250 to 0.
5. After the countdown, the device will be calibrated to 400ppm outside air.
6. Leave the CO2 monitor in the same place outdoors for at least another 15 minutes.
7. The CO2 monitor is now ready for use. The device does not need to be recalibrated
when it is turned off and on.
.