Input Calibration points
Range
Volts/Thermocouple
mA
RTD/Resistance
Individual calibration
Fine adjust
All
Input calibration
1. Save the current calibration to disk (select 'Save Calibration data'
from the 'Device' menu).
2. Select 'Calibrate Input' from the 'Device' menu.
. Enter the calibration password (Default is 101).
. Choose the calibration required (see Input calibration points table for
guidance).
5. Supply the signals requested by the software and follow the
instructions on screen. Press the 'Done' button when finished.
2
All manuals and user guides at all-guides.com
Values
Notes
10 mV, 20 mV,
If the internal switches are set to 10 V, you must supply
50 mV, 80 mV,
a signal ten times the displayed value, e.g. 10 V for the
200 mV, 1 V
1V point.
20 mA
100 Ω, 200 Ω,
Use four wire mode
500 Ω, 800 Ω,
2 kΩ, 10 kΩ
As selected
Allows you to recalibrate any of the points above
Current input
Fine tune the adjustment for the range you have selected.
All ranges
Complete calibration of all points. Internal switches must
be in the 1V position.