11. Attach Syringe 2 to the blood access site and slowly
draw the appropriate volume of blood. Gently mix with
ACD-A for thorough distribution. Refer to Table 1 for the
appropriate volumes of ACD-A and blood.
Note: Do not exceed 60 mL total volume.
Note: If multiple separation cycles are to be performed
with the same patient, repeat steps 10 and 11 to prepare
and fill syringes.
12. Disconnect Syringe 2 from the blood access site.
13. Remove the luer connector cap from the end of the
shorter length of chamber tubing and attach Syringe 1
(10-mL syringe, see Figure 7a).
14. Evacuate all air from Syringe 1 into the chamber tubing
line.
15. Place Syringe 1 into the appropriate syringe pump
receptacle.
16. Rotate and slide the plunger driver to engage the
Syringe 1 plunger (see Figure 7b).
17. Ensure whole blood is thoroughly mixed using gentle
agitation prior to placing in the syringe pump receptacle.
18. Remove the luer connector cap from the longer length of
chamber tubing and attach Syringe 2 (see Figure 7c).
19. Evacuate all air from Syringe 2 into the chamber tubing
line.
Note: Failure to remove air from syringes may
compromise the quality of the product.
20. Place Syringe 2 into the appropriate syringe pump
receptacle.
14
English