Figure 2. Thread the separation chamber tubing through the center
of the chamber caddy.
6.
Install the platelet separation chamber into the centrifuge
caddy (verify audible snap), make certain that both ends
of the chamber are properly located in the caddy notches
(see Figure 3a). Place the T-connector into the slot (see
Figure 3b), and tubing under the retainer on top of the
caddy surface (see Figure 3c). Press the tubing down into
the groove on the outer edge of the chamber caddy (see
Figure 3d).
Caution: Failure to install the separation chamber
properly may result in error codes.
Figure 3. Install the separation chamber into the caddy and place the tubing
into the slot and under the retainer.
English
11