Herunterladen Diese Seite drucken

Fluke 1550C Bedienungsanleitung Seite 49

Vorschau ausblenden Andere Handbücher für 1550C:

Werbung

1000 V
2500 V
5000 V
10,000 V
Bar graph range:
Insulation test voltage accuracy:
Induced ac mains current rejection:
Charging rate for capacitive load:
Discharge rate for capacitive load:
Leakage
Current
Measurement
Capacitive
Measurement
Timer
Live circuit
warning
< 200 kΩ
200 kΩ to 20 GΩ
20 GΩ to 200 GΩ
> 200 GΩ
< 200 kΩ
200 kΩ to 50 GΩ
50 GΩ to 500 GΩ
> 500 GΩ
< 200 kΩ
200 kΩ to 100 GΩ
100 GΩ to 1 TΩ
> 1 TΩ
< 200 kΩ
200 kΩ to 200 GΩ
200 GΩ to 2 TΩ
> 2 TΩ
Range
1 nA to 2 mA
0.01 μF to 20.00 μF
Range
0 to 99 minutes
Warning Range
30 V to 1100 V
ac/dc, 50/60 Hz
Insulation Tester
unspecified
5 %
20 %
unspecified
unspecified
5 %
20 %
unspecified
unspecified
5 %
20 %
unspecified
unspecified
5 %
20 %
unspecified
0 to 2 TΩ
-0 %, +10 % at 1 mA load current
2 mA maximum
5 seconds per μF
1.5 s/µF
Accuracy
±(20 % + 2 nA)
±(15 % of reading +
0.03 μF)
Resolution
Setting: 1 minute
Indication: 1 second
Voltage Accuracy
±(15 % + 2 V)
Specifications
31