Herunterladen Diese Seite drucken

Exchangeable Aperture - Altana BYK color2view Family Kurzbedienungsanleitung

Vorschau ausblenden Andere Handbücher für BYK color2view Family:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
BYK-Gardner GmbH

9 Exchangeable Aperture

The measurement aperture can be exchanged on color2view X models.
1 Aperture Plate 5 mm
2 Adjustment Ring 5 mm
Step 1: Select aperture in Configuration > Measurement Parameter > Aperture.
Step 2: Exchange the aperture.
• Use Plate Key (5) to manual change installed aperture plate.
• Attach Adjustment Ring (2 or 4) to Magnetic Lifter (6).
• Insert Magnetic Lifter and place it over the detector lens 32 mm in the middle.
• Press button on Magnetic Lifter's top to release Adjustment Ring.
• Check correct position of Adjustment Ring over the detector lens.
• Place new aperture plate on top and snap it into place.
• Confirm with check mark in upper right corner of aperture dialog.
2
1 Detector Lens 32 mm
2 20° Gloss Illumination
Step 3: After change instrument needs to be recalibrated, follow instructions on display.
If calibration fails, make sure that in the instrument display the correct aperture size was
selected.
301 500 209 - 2307
1
2
3 Aperture Plate 12 mm
4 Adjustment Ring 12 mm
32 mm Setup
1
3
4
5 Plate Key
6 Magnetic Lifter
12 mm Setup
3
2
3 20° Gloss Detection
4 Adjustment Ring 12 mm
color2view Family
5
6
4
3
13

Werbung

loading