Special safety notes
• This device has been designed for private use only and only for the purpose
provided.
• Keep the device outside the reach of children.
• Make sure that the measuring lens (6) is always clean.
• Put on the protective cap (1) after each use.
• Do not immerse the device in water.
• Do not let the device fall down and avoid heavy impacts.
• Do not try to disassemble or modify the measuring device.
• Do not use the device with if the measuring lens is damaged.
• In healthy persons, the temperature in different regions of the body may vary bet-
ween 0.2–1 °C (32.36 – 33.8 °F).
• The measuring accuracy of the device must be verified after 3 years. Please send
the device to your dealer or the next service point for this purpose.
Further notes
1. The measured body temperature may differ from one ear to the other. We recom-
mend to measure the temperature always at the same ear.
2. If you hold the device in your hand for more than 3 minutes this may lead to wrong
measuring results for any subsequent measurements. You should therefore avoid
to hold the device in your hand for too long.
3. If your child is younger than 2 years or if you do not measure your own temperature
(e.g., the temperature of older persons) we recommend to take three measure-
ments at the same ear. If the readings differ from each other choose the highest
temperature.
4. Minor deviations in repeated measurements are permissible.
5. If an obviously too low temperature should be measured the reason for this may be
found in the position of the lens.
6. If the device indicates a temperature that is too high it should not be used for about
30 minutes.
7. The temperature measured in the ear corresponds to the oral temperature.
8. This device is provided with an Anti Thermal Shock function in accordance with
EN 12470-5. This means that it works properly even in case of changes in ambien-
ce temperature (e.g., changes between rooms, changes between interior and
exterior temperature).
Safety symbols on the type plate
Carefully read the Operating Instructions before use
and keep them!
Application of Type BF
Description of the device
Denomination of the components
1. Protective cap
2. Display
3. On/Mem (MEMORY) button
4. Battery compartment
5. Sensor
6. Measuring lens
7. Measuring button
Putting in the batteries
The battery provided is a long-life lithium cell.
• Open the battery compartment (4).
• Replace the cell by a battery of the same type. Take care of the correct polarity.
• Close the battery compartment.