Setup
Make sure that the oxygen-sensitive surface is wetted safely and avoid air inclusion.
1. Secure the sensor to the stand.
2. Open the vial containing the zero-point gel.
3. Insert the sensor into the vial until it is approx. 10 to 15 mm from the bottom of the bottle.
 The sensor values change to the zero point.
4. Verify, calibrate or adjust the sensor. Typical times can be found in the table.
5. Then rinse the sensor in running water for approx. 30 seconds and dry.
 The sensor is now ready for use.
Fig. 4: Using the zero-point gel
Safety information:
This product is not classified as a hazardous material. Additional information can be found
on the MSDS safety data sheet. According to our experience and the information available
to us, this product is not damaging to health if used correctly and in the designated manner.
Disposal information:
The zero-point gel can be disposed of as regular household waste in accordance with the
official regulations.