Erstellt
von
Name:
Holger Seebode
Datum:
9.7.1998
#9
#10
#11
Response
0
Codes
5
3.7 HART-Command 12 : Read Message
Request
none
Data Bytes
Response
#0..#23 Message, Packed ASCII
Data Bytes
Response
0
Codes
5
3.8 HART-Command 13 : Read Tag, Descriptor, Date
Request
none
Data Bytes
Response
#0..#5
Data Bytes
#6..#17
#18..#20
Response
0
Codes
5
3.9 HART-Command 14 : Read Primary Variable Sensor Information
Request
none
Data Bytes
Response
#0..#2
Data Bytes
#3
#4..#7
#8..#11
#12..#15
Response
0
Codes
5
Bemerkung
HART-Protokoll
Gerät: COPA/MAG-XE, 50XE4000
Software: HART-Software
Kennung: D699B180 X.10
Projektnummer:
Device Identification Number, 24 Bit, MSB
Device Identification Number, 24 Bit
Device Identification Number, 24 Bit, LSB
No Command Specific Error
Incorrect Byte Count
No Command Specific Error
Incorrect Byte Count
Tag, Packed-ASCII
Descriptor, Packed-ASCII
Date: Day, Month, Year
No Command Specific Error
Incorrect Byte Count
Sensor Serial Number MSB, 24-bit unsigned integer
Sensor Limits/Min Span Units, Table II Unit Codes
Upper Sensor Limit, IEEE754
Lower Sensor Limit, IEEE754
Minimum Span, IEEE754
No Command Specific Error
Incorrect Byte Count
Sensor Serial Number
Upper Sensor Limit
Lower Sensor Limit
Minimum Span
Revision:
Name:
Datum:
= 0
= MSB Gerätenummer
= LSB Gerätenummer
= 0
= QmaxDN
= 0
= 0.02 oder 0.05 * QmaxDN,
(siehe unsigned char-Slot 28: Range <0.05RangeDN)
0
Holger Seebode
9.7.1998
Revision
Revision
Revision
3-3