Erstellt
von
Name:
Holger Seebode
Datum:
9.7.1998
3.10 HART-Command 15 : Read Primary Variable Output Information
Request
none
Data Bytes
Response
#0
Data Bytes
#1
#2
#3..#6
#7..#10
#11..#14
#15
#16
Response
0
Codes
5
Bemerkung
3.11 HART-Command 16 : Read Final Assembly Number
Request
none
Data Bytes
Response
#0..#2 Final Assembly Number
Data Bytes
Response
0
Codes
5
3.12 HART-Command 17 : Write Message
Request
#0..#23 Message, Packed-ASCII
Data Bytes
Response
#0..#23 Message, Packed-ASCII
Data Bytes
Response
0
Codes
5
HART-Protokoll
Gerät: COPA/MAG-XE, 50XE4000
Software: HART-Software
Kennung: D699B180 X.10
Projektnummer:
Alarm Select Code, Table VI
Primary Variable Transfer Function Code, Table III
Primary Variable Range Values Units Code, Table II
Primary Variable Upper Range Value, IEEE754
Primary Variable Lower Range Value, IEEE754, always Zero
Primary Variable Damping Value, IEEE754, Units of Seconds
Write Protect Code, Table VII
Private Label Distributor Code, Table VIII
No Command Specific Error
Incorrect Byte Count
Alarm Selection Code
PV Transfer Function Code
PV Upper Range Value
PV Lower Range Value
PV Damping Value
Write Protect Code
Private Label Distributor
No Command Specific Error
Incorrect Byte Count
No Command Specific Error
Incorrect Byte Count
Revision:
0
Name:
Holger Seebode
Datum:
9.7.1998
= 0 = High, 1 = Low
= 0 = Linear
= Qmax
= 0
= Dämpfung
= 251 = Not Implemented
= 18 =B-F&P
Revision
Revision
Revision
3-4