Herunterladen Inhalt Inhalt Diese Seite drucken

MACHEREY-NAGEL URYXXON Relax Bedienungsanleitung Seite 65

Vorschau ausblenden Andere Handbücher für URYXXON Relax:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Please make sure to remove access of urine from the test strip to avoid a carryover into the
instrument.
• Place the strip on the strip holder (Pic 3)
Pic 3: Correct
• Slide or push the strip to the end of the strip holder. Do not touch the reagent pads on the
test strip.
The instrument will automatically detect an applied strip. The measurement cycle will be
started. A progress bar on the display shows the remaining measurement time.
If "Autostart" (Chapter 9.5 "How to deactivate and activate the autostart") is deactivated,
the measurement must be started using the start control panel
The strip will be drawn into the instrument after 30 sec.
At the end of the measurement the result will be displayed on the screen and transferred to
the printer and interfaces.
By pressing the printer symbol
will lead back to the start screen.
Another analysis may be started by applying the next test strip.
CAUTION
NOTICE
NOTICE
RESULT
Seq.No: 001
ID: 1234567890
15.02.2019
9:47
*BLD
10 ERY/uL
UBG
NORM
BIL
NEG
PRO
NEG
NIT
NEG
KET
NEG
*GLU
50 mg/dL
pH
5.0
SG
1.000
LEU
NEG
Display 2: Result
the result can be printed again. Choosing the return panel
Manual URYXXON
Pic 4: Wrong
.
Relax EN, V 6.02 / 07.22
®
65

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis