Assemblies; Service Options; Calling The Service Menu - Richard Wolf Arthro 2202 Serviceanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

2

Assemblies

2.1

Service Options

2.1.1

Calling the service menu

SM--2202
Via the service menu you can change and display device parameters.
There are 4 menu items (L).
L0 Pressure Measurement
L1 Limit value calibration for pressure measurement
L2 Offset voltage of motor
L3 Speed calibration
.
IMPORTANT!
Always perform the menu items again after replacing an assembly.
Z Switch on the device.
Z While the actual pressure reading '-- -- --' blinks (1), touch the p
and p
buttons (2. Observe actuation sequence!). L1 appears on the
plus
nominal flow display. The 'L' blinks (3).
Z Push the start / stop button (4). The 'L' stops blinking. The nominal
pressure display (5) (and if applicable the actual pressure display) read
the current parameter of the menu level activated.
Z With the p
or the p
plus
the menu level.
Z Press the start / stop button. The 'L' starts blinking. The selected para-
meter is stored.
Z With the Q
or Q
plus
ted.
Z Switch off the device. After switching on again you may work with the
new parameters.
button (6) you can change the parameter of
minus
button (7) a different menu level can be selec-
minus
--
minus