Visual Check; Electrical Safety Check - Richard Wolf Arthro 2202 Serviceanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1.3

Visual check

Item to be checked
Device and accessories
Controls
Front plate and housing
Roller wheel
Membranes and covers
Labeling / Symbols
Safety--relevant inscriptions
(such as warnings)
Fuse inserts
Cabling
Printed circuit boards
Lamp plate (only light sources)
1.4

Electrical safety check

Item to be checked
Protective earth (PE) connection
Insulation resistance
Leakage current
SM--2202
The disinfectants used in the operating theatre as well as the ozone freed
when UV radiation sources are used, may change the surfaces of device
parts.
Check for
' Any soiling/contamination jeopadizing safety and general cleanliness
' Mechanical damage
' Loose or missing parts
' Mechanical function and easy operation
' Any damage having an effect on function and safety
' Function of control panels and LEDs
' Perfect condition, correct seating and easy operation
' Undamaged membranes of pressure sensors
' Undamaged cover of pressure chamber switch
' Completeness and good legibility
' Completeness and good legibility
' Values indicated by the manufacturer on the identification plate
(nominal current and melting characteristics)
' Perfect condition (position, insulation and possible brittleness)
' Corrosion or other damage
' Existence and good legibility
' Replace damaged parts immediately!
Check for
Measurement values:
' With supply cable not connected ± 0.1 Ohm
' With supply cable connected ± 0.2 Ohm
Test conditions:
10 A ± I
± 25 A , V
meas
0meas
Check:
The resistance between protective earth (PE) wire, protective earth (PE) con-
nection and the conductive parts connected.
Measurement values:
' Insulation resistance² 2 MOhm
Test conditions:
Test voltage 500V DC (continuous voltage)
Check:
All mains (power) cables relative to earth.
Measurement values:
' Device leakage current ± 100 µA
Test conditions:
Measuring setup (MD) in accordance with IEC 601
Check:
The leakage current which flows from each pole of the power supply unit via
the insulation through the protective earth (PE) wire or from the application
part via the patient to earth.
± 6 V , t
² 5 s
test
2