Guidance and Manufacturer's Declaration – Emissions All Equipment and Systems
The
®
Rapid Infuser, RI-2
Belmont
customer or user of the
Emissions Test
Compliance
RF Emissions
Group 1, Class A
CISPR 11
Harmonics IEC
Complies or Not
61000-3-2
applicable
Flicker
Complies or Not
IEC 61000-3-3
applicable
Guidance and Manufacturer's Declaration—Immunity All Equipment and Systems
Rapid Infuser, RI-2
The
®
Belmont
customer or user of the
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
IEC 61000-4-2
±8kV air
Radiated RF
3 V/m
IEC 61000-4-3
Electrical Fast
±2kV on AC Mains
Transient/burst
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Conducted RF
3 Vrms from
IEC 61000-4-6
150 kHz to 80 MHz
Power Frequency
3A/m
50/60Hz Magnetic
Field
IEC 61000-4-8
Voltage dips, short
>95% Dip for 0.5 Cycle
interruptions and
voltage variations
60% Dip for 5 Cycles
on power supply
input lines
30% Dip for 25 Cycles
IEC 61000-4-11
>95% Dip for5 Seconds
Kapitel 4: Parametereinstellungen und Vorbeugende Wartung
Table 201
is intended for use in the electromagnetic environment specified below. The
Rapid Infuser, RI-2
Belmont
®
The
Belmont
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Complies
Complies
Table 202
is intended for use in the electromagnetic environment specified below. The
Rapid Infuser, RI-2
Belmont
®
Compliance Level
±6kV contact
±8kV air
3 V/m
±2kV on AC Mains
±1kV Differential
±2kV Common
V1 = 3 Vrms
3A/m
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
should assure that it is used in such an environment.
Electromagnetic Enforcement – guidance
Rapid Infuser, RI-2
®
uses RF energy only for its internal
should assure that it is used in such an environment.
Electromagnetic Environment -
Floors should be wood, concrete or
ceramic tile. If floors are synthetic,
the relative humidity should be at
least 30%.
Radiated RF frequency is negligible
at any single frequency
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
If interference occurs, it may be
necessary to position the
Rapid Infuser, RI-2
sources of power frequency magnetic
field.
Power frequency magnetic fields
should be that of a typical location in
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Belmont
requires continued operation during
power mains interruptions, it is
recommended that the
Rapid Infuser, RI-2
an uninterruptible power supply or
battery.
43