Calibrating a measuring amplifier

7
Calibrating a measuring amplifier
1. Connect the USB power supply (use enclosed USB connection cable), PC
or external power supply to the K800.
USB‐Nr.
USB port for power supply/computer
communication
2. Connect the K800 instead of the S.G. quarter bridge to the MGCplus with
the 1-KAB263-3 connection cable (one cable for the AP814Bi, two cables
for the AP815i).
3. At the MGCplus, make the required settings for the calibration in accor­
dance with the Operating Manual.
4. Set the chosen detuning range at the step switch.
Set the step switch to 0% (the yellow status LED 0% lights up) and check the
zero point of the amplifier. Any variations from zero can be adjusted at the
MGCplus.
18
USBHBM1264
Socket for external 12 V power supply (customer
side, not included in scope of supply)
A01098_05_X00_00 HBM: public