Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration Check; Calibration Adjustment - Baker Hughes Druck UPS-III-IS Bedienungsanleitung

Eigensicherer stromschleifenkalibrator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25

Calibration Check

1.Connect the UPS-III-IS to the electrical calibrator. Switch on the electrical calibrator and allow it
to thermally stabilise.
2.Switch on the UPS-III-IS and allow the instrument to thermally stabilise.
3.Set the UPS-III-IS to mA measure, adjust the electrical calibrator to apply the first value in the
table 3. Record the reading of the UPS-III-IS.
4.Repeat step 3 for all the values in the table 3.
5.Compare the recorded values and the applied values if the difference is greater than the
permitted error, the instrument requires a calibration adjustment.
6.Repeat this procedure for V measure (table 4) and mA source (table 5).

Calibration Adjustment

INFORMATION: The instrument performs a two-point calibration for V and
mA.
1.Connect the UPS-III-IS to the electrical calibrator. Switch on the electrical calibrator and allow it
to thermally stabilise.
2.Switch on the UPS-III-IS and press
, within two seconds to select Calibration. Enter the
F2
access code [9410 factory setting] and allow the instrument to thermally stabilise.
3.Select the parameter required for calibration. Use the display menu to select the calibration
values. After a successful calibration enter the new calibration date.
2 seconds
Calibration menu
K0341 Revision B
8

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis