REAL TEMPERATURE LOGGING
Some models are equipped with reference containers
to make a real temperature logging.
If the unit is equipped with containers for real
temperature logging. It is important to prepare the
containers before use.
1.The containers is placed like shown in top fig.10 and
bottom fig.11
Fig.10
2.Remove the containers and fill it with 100ml with a
mixture of water and alcohol in ration 1:1 (if you are use
to use other products for reference containers you may
use this).
3.Place the containers back is the refrigerator and put
the sensor in the water. Make sure that all of the metal
on the sensor is covered in water.
OBS:
The mixture in the containers shall be checked
and refilled on a regular basis.
Fig.11
OPERATION AND FUNCTION
START UP:
When the appliance is connected to the power supply, the keyboard will automatically start up.
The start up view on the keyboard will show the different
software installed on the controller of the appliance.
Press enter to return from the start up view.
OPERATION – MAIN VIEW:
(1) Current temperature inside appliance (measured
by the TR3 probe)
(2) Time and date
(3) ALARM icon. Flashing by alarm. Remains visible after no active alarm until manually dis-
missed
(4) Logging icon. Visible when logging. Not visible
if no active logging.
(5) Memory icon. On when the memory is 90% full.
Flashing when the memory is full, and the controller
is deleting the oldest logging data.
(6) Memory bar. Shows the status of the memory.
KEYS – MAIN VIEW:
SET: Enter the Set Point menu of the temperature
SERVICE: Enter the Service menu
DATA: Enter the Data Logging menu
ALARM: Enter the Alarm menu
How to see and modify the temperature Set Point:
1. Push and immediately release the SET key: the display will show the Set Point value.
2. To modify the value push the SET key, the Set
Point starts flashing.
3. Use the UP and DOWN keys to modify the value.
4. To memorize the new Set Point value push the SET
key again or wait 30sec.