Table 206 – Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -
portable and mobile RF communications equipment and the Eva400EM/EE.
The Eva400EM/EE is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Eva400EM/EE can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Eva400EM/EE as
recommended below, according to the maximum output power of the communications equipment
Rated maximum
output of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
250
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
, 1
17
P
=
0,12
0,37
1,17
3,69
11,67
m
80 MHz to 800 MHz
d
, 1
17
P
=
0,12
0,37
1,17
3,69
11,67
800 MHz to 2,5 GHz
d
, 2
34
P
=
0,23
0,74
2,33
7,38
23,33