CYK10/11
7.4

Certificates and approvals

Declaration of Conformity
The product meets the requirements of harmonized European standards.It therefore meets the legal
specifications of EU guidelines.The manufacturer confirms that the product has been successfully tested
by applying the 4 mark.
7.4.1 Approvals for CYK10
IECEx
Ex ia IIC T3/T4/T6 Ga
ATEX
II 1G Ex ia IIC T3/T4/T6 Ga
FM / CSA
IS/NI CLI Div1&2 GP ABCD
Nepsi
II 1G Ex ia IIC T3/T4/T6 Ga
7.4.2 Temperature Classes
Cable
CYK10-G**a, a = 1, 2
CYK10-I**a, a = 1, 2
If above shown ambient temperatures are not exceeded there will be no invalid temperatures on
the simulator according to the temperature class.
Endress+Hauser
Measuring cable
-15 °C ... 135 °C
(5 °F ... 275 °F)
Range of T
for Temperature Class
a
T3
T4
-15 °C ... 120 °C
(5 °F ... 248 °F)
Identification
T6
-15 °C ... 70 °C
(5 °C ... 158 °F)