Wiring
CYK11 and fixed cable sensor
The colors of the cable match those of the sensors. Direct cabling is possible.
Some fixed cable sensors work with the Memosens operating voltage and are connected in the
same way as a Memosens sensor (e.g. CLS50D). Before commissioning check your sensor´s power
supply and connect the sensor accordingly.
CYK11 as extension for CYK10
The cores "GY" and "PK" are not required here. You must use a separate junction box or connection
panel to connect these cables. They must be terminated properly to avoid equipment damage!
The shields of both cables have to be connected. This is carried out automatically by the junction box
cable-to-cable through the shield contacting in the cable gland.
CYK11 with ferrule and M12 female connector
28
All manuals and user guides at all-guides.com
CYK10/11
a0016485-en
a0016486
a0015716-en
Endress+Hauser