Herunterladen Inhalt Inhalt Diese Seite drucken

VEGA VEGAPULS 62 Betriebsanleitung Seite 67

Füllstandsensor
Vorschau ausblenden Andere Handbücher für VEGAPULS 62:
Inhaltsverzeichnis

Werbung

primary_value
– Process value after min/max-adjustment and linearization. Selected as input to AIFB by setting
'Channel' = 1. Unit derives from 'Primary_value_unit'
primary_value_unit
– Selected unit code for "primary_value"
secondary_value_1
– Value after min/max-adjustment (level + level offset). Selected as input to AIFB by setting
'Channel' = 2. Unit derives from 'Secondary_value_1_unit'
secondary_value_1_unit
– Selected unit code for "secondary_value_1"
secondary_value_2
– Sensor value + sensor offset. Selected as input to AIFB by setting 'Channel' = 3. Unit derives
from 'Secondary_value_2_unit'
secondary_value_2_unit
– Unit code of 'Secondary_value_2'
sensor_value
– Raw sensor value, i.e. the uncalibrated measurement value from the sensor. Unit derives from
'Sensor_range.unit'
sensor_range
– 'Sensor_range.unit' refers to 'Sensor_value', 'Max/Min_peak_sensor_value', 'Cal_point_hi/lo'
simulate_primary_value
simulate_secondary_value_1
simulate_secondary_value_2
device_status
Linearization Type
– Possible types of linearization are: linear, user defined, cylindrical lying container, spherical
container
curve_points_1_10
– X and Y values for the user defined linearization curve
curve_points_11_20
– X and Y values for the user defined linearization curve
curve_points_21_30
– X and Y values for the user defined linearization curve
curve_points_31_33
– X and Y values for the user defined linearization curve
curve status
– Result of table plausibility check
SUB_DEVICE_NUMBER
SENSOR_ELEMENT_TYPE
display_source_selector
– Selects the type of value, which is displayed on the indication and adjustment module
max_peak_sensor_value
– Holds the maximum sensor value. Write access resets to current value. Unit derives from
'Sensor_range.unit'
min_peak_sensor_value
– Holds the minimum sensor value. Write access resets to current value. Unit derives from 'Sen-
sor_range.unit'
Calibration Highest Point
– Min./max.-adjustment: Upper calibrated point of the sensor. It refers to 'Cal_level_hi'. The unit
is defined in 'Sensor_range.unit'
Calibration Lowest Point
– Min./max.-adjustment: Lower calibrated point of the sensor. It refers to 'Cal_level_lo'. The unit
is defined in 'Sensor_range.unit'
VEGAPULS 62 • Foundation Fieldbus
10 Anhang
67

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis