STD GAS AUTO-CALIBRATION KIT (OPTIONAL)

When the STD gas auto-calibration kit MCO-SG is installed it is possible to calibrate CO
manually.
1. Connect a standard gas cylinder to the
connecting port on the lower right side of the
incubator. Use a cylinder with the same CO
O
density as the density set on the incubator.
2
Note: There is no problem in keeping the standard
gas cylinder connected after CO
density calibration is completed.
2. Go to the Tools #1 screen (Menu
Press "STD Gas Setting" to display the STD Gas
Setting screen.
3. Enter the CO
connected standard gas cylinder. Press "Apply".
The display returns to the Tools #1 screen.
CO
settable range: 4.0 %–21.0 %; factory setting:
2
5.0 % (ON). If not necessary, slide to OFF.
O
settable
range:
2
22.0 %–81.0 %. Factory setting: 5.0 % (ON). If not
necessary, slide to OFF.
4. Press "STD Gas Calibration" to display the STD
Gas Calibration screen.
68
and/or
2
and/or O
2
Tools#1).
and/or O
density of the
2
2
0.5
%–18.0
%
Lower right side
gas
2
Connecting port for standard gas cylinder
and
and/or O
density
2
2