4. Operation
Analysis Result
If CardiAid cannot perform a clear heart rhythm analysis due to incorrect electrode connection
or inadequate signal level which is a result of touching or moving the patient, the warning
"Movement detected. Do not touch the patient from now on." is heard. CardiAid
will attempt another heart rhythm analysis. If the analysis is successful, device proceeds according
to the results as explained in sections "Shock Necessary" or "Shock Not Advised". If the
second attempt is also unsuccessful, the device proceeds to "Basic Life Support" phase.
Signs of Life Present
If you detect signs of life like consciousness and normal breathing during operation of the device,
place the patient in recovery position (lying on one side). Do not remove electrodes. CardiAid will
continue operating normally. Heart rhythm analysis will be performed regularly.
If the patient loses consciousness again and "Shock necessary" instruction is heard, place the
patient on his back again and follow CardiAid's instructions.

Information Function

During operation of CardiAid, information about the duration of use and the number of shocks
delivered may be obtained. Press info-button to obtain this information. Information can only be
provided while electrodes are not connected or the basic life support is being performed. The info-
button is not active during heart rhythm analysis and shock delivery phase. If info-button is pressed at
these stages, the information will be provided when CardiAid proceeds to basic life support phase, or
the electrodes are disconnected.

4.3 After Using CardiAid

• Remove the plug of the electrodes from the socket.
• Close the cover of the device.
• Contact Cardia International or a Cardia International authorized service provider immediately
after use.
• Remember that electrodes and the emergency kit are disposable and should be replaced after
use.

4.4 Operation Documentation

Operation Data
The following information is saved automatically in the internal memory of the device for every use:
• Date and time of use
• Patient's ECG
• Time of each voice prompt
• Time of important points in resuscitation like when the device starts and finishes the analysis,
what the result of the analysis is and when the shock-button is pressed
• Time and number of shocks delivered
Contact Cardia International or a Cardia International authorized service provider immediately after
use.
20