Applying the overtape
1
1. Remove the adhesive
liner marked "1" from
the overtape. Do not
remove the adhesive
liner marked "2"
from the sides of the
overtape.
This image shows an example of the Enlite overtape applied correctly. To apply the
overtape, go to step 1.
Note: The sensor body and the skin must be covered partially as shown in
the image. The sensor connector and sensor snaps must be uncovered and
accessible.
2
2. Apply the overtape
to the rounded part
of the sensor and
the skin, making
sure that the sensor
connector and sensor
snaps are sticking out
through the hole in the
overtape.
3
3. Stretch the overtape
around the sensor
connector and sensor
snaps so that it lies
flat against the sensor
adhesive pad.
- 14 -
4
4. Remove the adhesive
liner marked "2" from
both sides of the
overtape and smooth
the overtape to the
skin.