Herunterladen Inhalt Inhalt Diese Seite drucken

Connection Of The High Vacuum Flange - Agilent TwisTorr 305-ICQ Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Instructions for Use

Connection of the High Vacuum Flange

If the rotor stops suddenly, the resulting torque must be absorbed by the hardware
fastening the high vacuum flange.
Bolts or hardware must meet the specification below.
It is mandatory to connect the pump in such a way to withstand the torque
WARNING
specified in the "INSTRUCTION FOR USE" section. Specifically pay
attention to clamp design, material of clamps and bolts and bolt fixing
torque.
Use only high resistance class clamps (or bolts) ≥ 8.8.
If the pump is fastened on a vacuum chamber with flanges different
WARNING
from those recommended by Agilent, there is the chance of the
TwisTorr 305-ICQ & 305-IC SF twisting or tearing-off in case the rotor
suddenly stops. The combination of different flange types is not allowed.
Agilent will not accept any liability for all damages resulting from improper
fastening.
190
Agilent TwisTorr 305-ICQ & 305-IC SF User Manual

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Twistorr 305-ic sf

Inhaltsverzeichnis