Function Check with the
ReproEasy Pad
Quick Reference Guide for Sartorius Models MA37|160
The performance test on models MA37|160 enables you to monitor
your device's function, testing both the heating unit and the
weighing system at the same time at the temperature you selected.
1. Start the performance test
– Select "MA Performance Test" under the "Calibration/Adjustment"
item in the Setup menu.
– Select a temperature close to the one you use from the list of
temperatures suggested.
2. The device now shows you the next work steps
3. Place the reference sample on the sample pan
4. Place the sample pan with the reference sample
"ReproEasy Pad" on the pan retainer.
5. Close the Hood
– The performance test starts automatically.
Evaluation of performance test
The device is in a perfect running state and displays "Device is OK"
if the Moisture Analyzer performance index (MApi) value falls
within the following ranges:
From serial no.: 00343xxxxx
Temperature adjustment
85 °C
125 °C
165 °C
Up to and including serial no.: 00342xxxxx
Temperature adjustment
70 °C
100 °C
130 °C
Due to the effect temperature has on the results, the test is only
possible once the device has cooled down. If the device is too warm,
a relevant message is displayed:
Contact the Sartorius Service Center if the
message „Device is not OK." appears.
Sartorius Lab Instruments GmbH & Co. KG
Otto-Brenner-Strasse 20
37079 Goettingen, Germany
Phone +49 (0)551 308 0
www.sartorius.com
Specifications subject to change without notice.
Copyright Sartorius Lab Instruments GmbH & Co. KG.
Printed in the EU on paper bleached without chlorine.
Publication No.: WMA6070-a161102 · Ver. 11| 2016
– Open the packaging of the ReproEasy Pad
and remove the pad.
– Remove the protective film.
– Affix the pad to the center of the sample pan,
sticky side down.
MA performance test
The device is too warm for a performance
test.
The device is still cooling down.
m
Tolerance range
between 9 and 11 MApi
between 16 and 18 MApi
between 21 and 25 MApi
Tolerance range
between 9 and 11 MApi
between 16 and 18 MApi
between 21 and 25 MApi

Werbung

loading

Inhaltszusammenfassung für Sartorius ReproEasy Pad MA37

  • Seite 1 Otto-Brenner-Strasse 20 37079 Goettingen, Germany Phone +49 (0)551 308 0 www.sartorius.com Specifications subject to change without notice. Copyright Sartorius Lab Instruments GmbH & Co. KG. Printed in the EU on paper bleached without chlorine. Publication No.: WMA6070-a161102 · Ver. 11| 2016...