I I n n s s t t a a l l l l a a t t i i o o n n c c h h e e c c k k l l i i s s t t
Follow this checklist for the 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t :
Confirm that you do not have any unused components after installation. Your S S e e c c u u r r e e C C o o n n n n e e c c t t does not ship with
any extra components
Check that the S S e e c c u u r r e e C C o o n n n n e e c c t t ID number and room number/location has been recorded on the S S e e c c u u r r e e C C o o n n n n e e c c t t
association form (page 37)
Use a tape measure to check that the S S e e c c u u r r e e C C o o n n n n e e c c t t is installed at the horizontal center of the wall behind the bed
location
Use a tape measure to check that the bottom of the S S e e c c u u r r e e C C o o n n n n e e c c t t is installed 4 in. - 5 in. (10.2 cm - 12.7 cm) from
the floor
Use a level to confirm that the mounting plate is level
All fasteners are tight with no signs of protruding or missing fasteners
Power supply is plugged into a hospital grade protective earthed wall outlet and to the S S e e c c u u r r e e C C o o n n n n e e c c t t
Nurse call communication cable is plugged into the S S e e c c u u r r e e C C o o n n n n e e c c t t and the nurse call system
Product ID number:
Installed by:
Inspected by:
N N o o t t e e - - Maintain a copy of this record for at least 10 years.
EN
8
Date:
Date:
5212-009-205 Rev AA.1