25
Recommended separation distance between portable and mobile radio-communication devices and the monitor
The SUPER VEGA BATTERY SUCTION UNIT surgical aspirator is intended to operate in an electro-magnetic envi-
ronment where RF irradiated interferences are under control. The client or operator of the SUPER VEGA BATTERY
SUCTION UNIT device can help prevent electro-magnetic interference by keeping a minimum distance between the
portable and mobile RF communication devices (transmitters) and the SUPER VEGA BATTERY SUCTION UNIT device,
as recommended below, in relation to the radio-communication maximum output power.
Maximum nominal
output power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people
SYMBOLS
Caution: read instructions (warnings) carefully
Keep in a cool, dry place
Manufacturer
Product code
Medical Device complies
with Directive 93/42/EEC
WEEE disposal
Serial number
Battery
Direct current
IP21
Covering Protection rate
ON / OFF
Separation distance from the frequency transmitter (m)
150KHz to 80MHz
3,5
√P
d=
V
1
0,12
0,38
1,2
3,8
12
ENGLISH
80MHz to 800MHz
12
√P
d=
E
1
0,12
0,38
1,2
3,8
12
Follow instructions for use
Keep away from sunlight
Date of manufacture
Lot number
Type BF applied part
Class II applied
Temperature limit
Atmospheric Pressure limit
~
Alternating current
Hz
Mains frequency
Humidity limit
800MHz to 2,7GHz
23
√P
d=
E
1
0,23
0,73
2,3
7,3
23