GuRecommended separation distance between portable and mobile RF communications equipment and the iNAP One
The iNAP One is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the iNAP One can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmit-ters) and the iNAP One as recommended below,
according to the maximum output power of the com-munications equipment.
IRated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
d =1,2√P
d =1,2√P
0,12
0,12
0,38
0,38
1,2
1,2
3,8
3,8
12
12
800 MHz to 2,5 GHz
d =2,3√P
0,23
0,73
2,3
7,3
23
44

Werbung