Guidance and manufacturer's declaration-electromagnetic immunity
The iNAP One is intended for use in the electromagnetic environment specified below.
The customer or the user of the iNAP One should assure that is used in such and environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 KHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80MHz to 2,5 GHz
Compliance level
Electromagnetic environment-guidance
3 Vrms
Portable and mobile RF communications equipment
should be used no closer to any part of the iNAP One
including cables, than the recommended separation
3 V/m
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).