Guidance and manufacturer's declaration-electromagnetic immunity
The iNAP One is intended for use in the electromagnetic environment specified below.
The customer or the user of the iNAP One should assure that it is used in such an environment.
Immunity test
Voltage Dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency(50, 60 Hz)
magnetic field IEC 61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
41
IEC 60601 test level
Compliance level
<5% UT(>95% dip in
<5% UT(>95% dip in
UT) for 0,5 cycle
UT) for 0,5 cycle
40% UT(60% dip in
40% UT(60% dip in
UT) for 5 cycles
UT) for 5 cycles
70% UT(30% dip in
70% UT(30% dip in
UT) for 25 cycles
UT) for 25 cycles
<5% UT(>95% dip in
<5% UT(>95% dip in
UT) for 5 s
UT) for 5 s
3 A/m
3 A/m
Electromagnetic environment-guidance
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the iNAP One
requires continued operation during
power mains interruptions, it is
recommended that the iNAP One be
powered from an uninterruptible power
supply or a battery.
The iNAP One power frequency magnetic
fields should be at levels characteristic of
a typical location in a typical commercial
or hospital environment.