Guidance and manufacturer's declaration-electromagnetic immunity
The iNAP One is intended for use in the electromagnetic environment specified below.
The customer or the user of the iNAP One should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge(ESD) IEC 61000-
4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge IEC 61000-4-5
IEC 60601 test level
Compliance level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
± 2kV for power
± 2kV for power supply
supply lines
lines
± 1kV for input/
Not applicable
output lines
± 1kV line(s) to line(s)
± 1kV differential
± 2kV line(s) to earth
mode
Not applicable
Electromagnetic environment-guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
The iNAP One power frequency magnetic
fields should be at levels characteristic of
a typical location in a typical commercial
40