Explanation of symbols

Caution: read
instructions (warnings)
carefully
WEEE disposal
Product code
Manufacturer
Temperature limit
Covering Protection
IP21
rate

Electromagnetic Compatibility Information

The device satisfies the EMC requirements of the international standard IEC
60601-1-2. The requirements are satisfied under the conditions described in the
table below. The device is an electrical medical product and is subject to special
precautionary measures with regard to EMC which must be published in the in-
structions for use. Portable and mobile HF communications equipment can affect
the device. Use of the unit in conjunction with non-approved accessories can af-
fect the device negatively and alter the electromagnetic compatibility. The device
should not be used directly adjacent to or between other electrical equipment.
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emissions test
Radiated emission
CISPR 11
Conducted emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
ENGLISH
Compliance
Group 1,
class B.
Group 1,
class B.
Class A
Complies
Authorized represent-
ative in the European
community
Keep away from
sunlight
Lot number
Date of manufacture
Follow instructions
for use
Electromagnetic environment - guidance
The device uses RF energy only for its internal
function. Therefore, its emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establish-
ments, including domestic establishments and
those directly connected to the public low-volt-
age power supply network that supplies build-
ings used for domestic purposes.
Medical Device com-
plies with Directive
93/42/EEC
Serial number
Direct Current
Humidity limit
Type BF applied part