switches operating modes. If you are adjusting a value you will
lose those changes.
Õ
toggles between increment and decrement mode.
Ù
â
,
Calibrate/Run mode selection jumper
Place the calibration jumper in the CAL position to start calibrating. You
will have to open the case to select the jumper and normal precautions
against ESD are essential. Take careful note of the position of the wires
connecting the front and back of the case together.
Voltage Source Mode
Display
Switch on Power, Calibrate link In CAL Position, Measure output voltage
u=0
u=0
u=10
u=10
Sure
Current Source Mode
Display
Switch on Power, Calibrate link In CAL Position
u=0
Connect multimeter to measure output current (mA).
6
QS61001066_Portacal1000_Buch.ind26 26
á
and
adjust the output,
I+
I-
V
Off
Off
On
I+
I-
V
Off
Off
On
á
provides the finest adjustment.
Keys
Action
Adjust Output Voltage to 0 V
Õ Ù â á
Accept output
Ó
Adjust Output Voltage to 10 V
Õ Ù â á
Accept output
Ó
To Save Values (or any other key to restart).
Ó
Keys
Action
Selects Current Source Mode
Ò
25.01.2008 14:52:54 Uhr