NOTE
Pre‐stress forces must act precisely in the measurement direction. Bending
moments or lateral forces must not occur. Otherwise the transducer can be
destroyed.
Force to be measured
Ground, Ra0.5
Fig. 5.2: Force washer pre‐stressed by screw
WARNING
If you will be introducing tensile forces, observe the maximum loading.
There is a danger of breakage if permissible loading is exceeded. This
can cause danger for the operating personnel of the system in which the
transducer is installed.
When tensile forces are exerted, the tensile force acts in addition to the
pre‐stress force on the pre‐stressing elements. These elements must
therefore be designed appropriately.
Mounting is as shown in Fig. 5.2. Adjust the pre‐stress by measuring the force
washer. The recommended values for pre‐stress can be found in the following
table depending on the CLP type.