G1 - G9
3.
Connect the Air inlet port of the generator (option 0 only) to a suitable compressed air supply using either copper or nylon tubing.
The air quality must meet the requirements of ISO 8573.1:2001, Class 2. -. 1 and should not exceed 9 Bar.
4.
The Mist-X exhaust should be connected to the Drain outlet port (G4 Models only). It is recommended that the silencer be placed in
a suitable container or drain so that the mist is contained safely.
MIST-X
5.
All piping and components external to the generator should be suitable for the application and rated to at least the maximum operating
pressure of the generator. Copper or stainless steel pipes should be used to connect the gas outlets to the application instrument and
sized in accordance with table 3.3. These pipes should be adequately supported to prevent damage and leaks in the system.
Installation Kits
Description
1/8" N
LAB Fittings Basic
2
1
Accessories kit.
1/4" N
LAB Fittings Basic Accessories
2
2
kit.
3.3

Electrical Installation

Attach the appropriate mains supply cable provided to the IEC 320 socket located on the rear of the generator and connect to the
electrical supply.
If a cordset, other than the one provided, is used to connect the equipment to the electrical supply ensure that it is suitably rated for
the application and in accordance with local and national code regulations.
The equipment must be bonded to earth through the cordset.
Warning
dh Part No:
606272372
606272378
9
Contents
Ball. valve 1/4" BSP (x1)
1/4" Non return valve (x1)
Reducer 1/4" - 1/8" BSP (x4)
Adaptor 1/8" BSP M 1/8" OD (x4)
Ball. valve 1/4" BSP (x1)
1/4" Non return valve (x1)
Reducer 1/4" - 1/8" BSP (x4)
Adaptor 1/8" BSP M 1/4" OD (x4)