Operation
d. Go to the router's configuration page via a browser by
entering the IP address 192.168.10.100:8080.
e. Use the standard login credentials:
Username: adm
Password: 123456
f. Then enter the specified IP address and the gateway
via the menu item Network -> WAN. (The MAC address
can also be read in this mask if it is required.)
g. The Air Guard can now be connected to the network
and all other network functions can be set up.
NOTICE
In order that the PC used can be reconnected to the network,
the TCP / IPv4 address must be changed to "obtain
automatically" or to the previous settings.
4.4.8

Configuration Mail Server

SMTP server Enter the IP address of the sending mail server
(SMTP).
Sender
Enter the sender address.
address
Email
User name for logging in to send emails
Username
Email
Password for logging in to send emails
Password
Ready
The Ready display indicates whether the
entered values are valid. The control verifies
the values after the user taps the Set config
button.
Green - Configuration valid
Grey - Configuration invalid
Test-Email
Sends a test email to the email address
configured in the receiver area (see chapter
4.4.9)
Write
The user must tap the Set config button to save
configuration
the settings.
22
4.4.9

Configuration Email Receiver

Receivers
Enter the email addresses of the receivers. Up
1-5
to five receivers can be entered.
ON/OFF
Activate or deactivate an email address.

4.4.10 Configuration Remote Control

1. The SmartClient from Siemens is required in order to be
able to use the Air Guard's remote control. The file
"SmartClient.exe" can be downloaded from the Siemens
homepage.
2. After downloading the application, it can be started without
installation.
3. After starting the application, the "external IP address"
stored in the Air Guard must be entered (see chapter
4.4.6).
4. Then enter the password for access to the Air Guard. On
delivery, this is always "airguard". The password can only
be changed by Dräger Service or by Dräger authorised
technical personnel, since special access rights are
required for this.
5. After a successful connection, a window showing the Air
Guard display appears. The keys F1 to F4 correspond to
those below the touchscreen (see chapter 4.4).
6. Using the mouse and keyboard, all entries can now be
made that can also be made directly on the Air Guard's
touchscreen.
NOTICE
While entries are made via the SmartClient, a warning triangle
appears on the Air Guard's touchscreen, indicating the remote
control. Conversely, it is identical if entries are made on the Air
Guard's touchscreen.
CAUTION
Risk of injury due to unexpected activation of the purge
The purge of the Air Guard, too, can be activated via the
remote control. Purging can occur very unexpectedly for
persons in the vicinity of the Air Guard.
► The operator must introduce suitable preventive measures
so that there are no dangers due to the sudden purging
(e.g. instructing staff on the features of the Air Guard and
wearing hearing protection when working near the Air
Guard).
► Observe national and regional regulations and laws.
CAUTION
Risk of injury if hearing protection is not worn
The gas is released at a very high speed during purging and is
extremely loud.
► Hearing protection must be worn during purging.